Update readme.md
This commit is contained in:
parent
32c2db5838
commit
cb47ee62ff
20
readme.md
20
readme.md
@ -1,5 +1,19 @@
|
|||||||
## R6581T Adjustment Toolkit
|
## R6581T Adjustment Toolkit
|
||||||
A toolkit to dump INL calibration constants of the R6581 DMM and update them to fit a reference DMM's.
|
A toolkit to dump INL calibration constants of the R6581 DMM and update them to fit a reference DMM's.
|
||||||
- Run toolkit with two meters
|

|
||||||
- Compensate INL
|
|
||||||
- Look at data
|
# How to set up
|
||||||
|
1. Download repository
|
||||||
|
2. Open Hosei.py and adjust the following for your setup:
|
||||||
|
- libraries (uses a pyvisa compliant alternative for USB-CDC GPIB adapter)
|
||||||
|
- lines #19-21 for the GPIB adresses
|
||||||
|
- lines #36-42 for your voltage source configuration
|
||||||
|
- lines #43-54 for your reference dmm configurations
|
||||||
|
- the measurement functions from lines #74 to 92 for your DMM and source.
|
||||||
|
3. You can opt to comment out line #63 to 72 if you do not wish to clear your initial coefficients for the first sweep.
|
||||||
|
4. Run and wait
|
||||||
|
|
||||||
|
There are some basic pareametersd also like nrepeats, which defines the amount of INL sweeps to be averaged.
|
||||||
|
Note that this is a tradeoff of thermal effects against DMM noise
|
||||||
|
|
||||||
|
Typically with Datron 4805 and 3458A I get within +-0.1 uV/V of the reference DMM (3458A) on an average correction fit.
|
||||||
Loading…
x
Reference in New Issue
Block a user